Mounting And Installation; Sensor Enclosure - E+E Elektronik HTS801 User Manual

Humidity and temperature sensor for high humidity and chemically polluted conditions
Table of Contents

Advertisement

Type T17 Probe
Pressure-tight up to 20 bar (300 psi)
with optional cut-in fitting
RH
code "probe length"
T
code "cable length"
150 (5.9)

4 Mounting and Installation

4.1

Sensor Enclosure

Mounting with Screws
▪.
Drill the mounting holes according to the corresponding mounting pattern below.
▪.
Mount the back section of the enclosure with 4 screws (screw diameter < 4.2 mm (0.2"), not included in the
scope of supply).
Values in mm (inch).
7 (0.28)
Fig. 25
Drilling pattern of polycarbonate enclosure
1)
code "cable length"
1)
1/2".ISO.or.1/2".NPT
1)
1/2".ISO.or.1/2".NPT
166 (6.54)
5.1 (0.20)
Type T28 Probe
For meteorological applications
RH
39
93 (3.66)
(1.54)
T
code "cable length"
150 (5.9)
200 (7.87)
7 (0.28)
Fig. 26
Drilling pattern of stainless steel enclosure
Humidity and Temperature Sensor for High Humidity and Chemically Polluted Conditions | 23
User Manual HTS801
1)
code "cable length"
53 (2.09)
1)
35 (1.38)
5.1 (0.20)