Setting Up The Ni Cdaq-9172 - National Instruments NI-9203 User Manual And Specifications

Hide thumbs Also See for NI-9203:
Table of Contents

Advertisement

Setting Up the NI cDAQ-9172

Note
The NI-DAQmx software is included on the CD shipped with your kit and is
available for download at
after installation from Start»All Programs»National Instruments»NI-DAQ»DAQ
Getting Started Guide. Other NI documentation is available from
Note
Additionally, attach a wire with a ring lug to all other C Series I/O module cable
shields. You must connect this wire to the ground terminal of the chassis using the ground
screw.
© National Instruments Corporation
Complete the following steps to prepare the NI cDAQ-9172 chassis for use:
1.
Before connecting the hardware, install the NI-DAQmx software.
Refer to the DAQ Getting Started Guide for more information about
software installation.
ni.com/support
2.
If you are not using any mounting accessories, attach the provided
rubber standoffs to the back of the NI cDAQ-9172 chassis.
3.
Make sure the NI cDAQ-9172 chassis power switch is turned off.
4.
Attach a ring lug to a 14 AWG (1.6 mm) wire. Connect the ring lug to
the ground terminal on the side of the chassis using the ground screw.
Attach the other end of the wire to the system safety ground.
1 Attached to System Ground
Figure 6. Ring Lug Attached to Ground Terminal
5.
Remove the plastic cover from the connector in any empty module slot.
. The DAQ Getting Started Guide is available
15
NI cDAQ-9172 User Guide and Specifications
ni.com/manuals
1
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ni cdaq-9172