NI cDAQ-9172 User Guide and Specifications
You can route the signal to measure to the Source input of Counter 0,
as shown in Figure 27. Assume this signal to measure has frequency F1.
Configure Counter 0 to generate a single pulse that is the width of N periods
of the source input signal.
Signal to
Measure (F1)
Signal of Known
Frequency (F2)
CTR_0_SOURCE
(Signal to Measure)
CTR_0_OUT
(CTR_1_GATE)
CTR_1_SOURCE
Figure 27. Frequency Measurement—Method 3
Then route the Counter 0 Internal Output signal to the Gate input of
Counter 1. You can route a signal of known frequency (F2) to the Counter 1
Source input. F2 can be 80MHzTimebase. For signals that might be slower
than 0.02 Hz, use a slower known timebase. Configure Counter 1 to
perform a single pulse-width measurement. Suppose the result is that the
pulse width is J periods of the F2 clock.
From Counter 0, the length of the pulse is N/F1. From Counter 1, the length
of the same pulse is J/F2. Therefore, the frequency of F1 is given by
F1 = F2 * (N/J).
SOURCE
COUNTER 0
SOURCE
GATE
0
1 2
3 ... N
Interval to
Measure
52
OUT
OUT
COUNTER 1
ni.com