Semi-Period Measurement

In semi-period measurements, the counter measures a semi-period on its
Gate input signal after the counter is armed. A semi-period is the time
between any two consecutive edges on the Gate input.
You can route an internal or external periodic clock signal (with a
known period) to the Source input of the counter. The counter counts
the number of rising (or falling) edges occurring on the Source input
between two edges of the Gate signal.
You can calculate the semi-period of the Gate input by multiplying the
period of the Source signal by the number of edges returned by the counter.
Single Semi-Period Measurement
Single semi-period measurement is equivalent to single pulse-width
measurement.
Buffered Semi-Period Measurement
In buffered semi-period measurement, on each edge of the Gate signal, the
counter stores the count in a hardware save register. The NI cDAQ-9172
transfers the stored values to host memory.
The counter begins counting when it is armed. The arm usually occurs
between edges on the Gate input, which means that the first value stored in
the hardware save register does not reflect a full semi-period of the Gate
input. In most applications, this first point should be discarded.
Figure 23 shows an example of a buffered semi-period measurement.
Counter Armed
GATE
SOURCE
Counter Value
0
Buffer
Figure 23. Buffered Semi-Period Measurement
1
2
1
2
2
2
48
3
1
1
1
2
2
2
2
3
1
2
3
3
3
1
1
2
ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ni cdaq-9172

Table of Contents