Reloading the
If the instrument does not respond to commands or responds
inappropriately, the firmware image may be corrupted. You can reset the
Firmware
firmware by performing a total reset (also known as a double reset).
Performing a Total Reset
A total reset erases the current firmware image from instrument
memory. This is indicated by the instrument status lights changing from
green (ready) to flashing yellow.
To reset the firmware image with a total reset:
Step
After a total reset, a new copy of the firmware will be downloaded to the
instrument when you relaunch the ABI P
software.
3-6 Software and Firmware
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Action
1
Exit Collection.
2
Press the reset button on the back of the 377 instrument.
3
Immediately press the reset button again.
377-96 Collection
RISM