Programming mode

In direct access mode, pressing PRG enters programming mode, where the unit's main operating parameters can be set.
To access the Service parameters, go to the "PSD" menu item (see the table below) and enter the password. The user
terminal only provides access to the basic confi guration parameters, such as direct functions and active alarms without
password protection, or, with password protection (Service).
Notice: in the default settings the user password is set to 0 and is not required when entering programming mode; if the
password has been set to a value other than 0, this needs to be entered to access programming mode. User will still have
access to the Direct Access Mode without entering a password.
The menu items available and parameters visible on the user terminal are listed below.
Table 9. Programming Menu
dir (Direct
nnt
functions)
See the
rL2 (**)
following table
rH2 (**)
rt (**)
ESC

Temperature settings

The factory default temperature setting is 5°C for all laboratory and pharmacy refrigerators. To change the factory settings,
refer to programming instructions.
The refrigerator described in this manual is designed for optimum performance at 5°C. It is advisable to call Technical
Service before changing set points.
The recommended set point for NSF 456 vaccine refrigerators is 5°C. Changing this parameter or any other service
parameters may aff ect the units ability to meet the NSF 456 standard.

Direct functions

Table 10. Direct Function Parameters
Acr.
Description
boF
Activate battery storage status
btr
Start battery test
dFn
Start defrost
Eco
Activate Eco mode
Fr
Firmware version
OnF
Unit ON/OFF
SAh
Display alarm log
Sd
Defrost probe
SPr
Product probe
SrG
Control probe
St
Set control set point
CtL
Pro
(Control)
(Display
probes)
St
/ 5
Sth
/ cr
HU
/ cc
ESC
ESC
HcP
CnF
(Temperature
(Confi guration)
alarms)
HAn
Hb
HFn
PDU
rHP
ESC
ESC
ALn
PSD ESD
(Alarms)
AHA
ALA
Ad
Add
rSA
ESC
25