Step 5. Performing an operational check

Step 5. Performing an operational check

Improper connector care, cleaning, or use of mismatched cable connectors can
invalidate the published specifications and damage connectors. Clean all cables
before applying to any connector. Repair of damaged connectors due to
improper use is not covered under warranty.
for Accurate Measurements" on page 1-12
Return loss can be tested using a number of different test equipment configu-
rations. Some of these are:
receiver and lightwave coupler
Many other possibilities exist. The basic requirements are an appropriate
lightwave source, a compatible lightwave receiver, and a compatible lightwave
coupler.
Refer to the manuals provided with your lightwave test equipment for infor-
mation on how to perform a return loss test.
Typical return loss is better than 40 dB. For actual specifications on your par-
ticular cable or accessory, refer to the manufacturer.
Insertion loss can be tested using a number of different test equipment config-
urations. Some of these are:
a lightwave source and receivers
ule
2-8
Refer to "Cleaning Connections
for proper cleaning procedures.