Agilent Technologies 86060A User Manual page 83

Lightwave switches
Table of Contents

Advertisement

Listing
Description
Programming
Programming Examples
Example 1: Switch position using the *WAI command
This program prompts the operator for the desired switch position and then
moves the switch to this position. The switch error queue is then read and
printed. The program shows how to use the *WAI command to ensure that the
switch has settled to its new position.
10
INTEGER Switch_addr,A_position,B_position
20
DIM Command$[80],Channel$[80],Error_return$[80],
Dummy$[80]
30
Switch_addr=711
40
CLEAR SCREEN
50
INPUT "Enter A-port position : ",A_position
60
INPUT "Enter B-port position : ",B_position
70
Channel$="A"&TRIM$(VAL$(A_position))&",B"&TRIM$
(VAL$(B_position))
80
Command$="ROUTE:LAYER1:CHANNEL "&Channel$
90
OUTPUT Switch_addr;Command$
100
GOSUB Wait_to_settle
110
REPEAT
120
OUTPUT Switch_addr;"SYSTEM:ERROR?"
130
ENTER Switch_addr;Error_return$
140
PRINT Error_return$
150
UNTIL (VAL(Error_return$)=0)
170
GOTO Exit_prog
180
!
190 Wait_to_settle: ! wait for switch to settle
200
OUTPUT Switch_addr;"*WAI"
210
OUTPUT Switch_addr;"SYSTEM CONFIG?"
220
ENTER Switch_addr;Dummy$
230
RETURN
240
!
250 Exit_prog:!
260
END
Line No. 1
10 to 20
Declare some variables for use in the program.
30
Set the Agilent 8606X Optical Switch address variable,
Switch_addr, to 711 (factory default).
40 to 60
Clear the screen and prompt the operator for the desired
switch position. Store the positions in variables
A_position and B_position.
70
Set Channel$ to represent the switch channel positions in
4-32