EDAN Norav NECG-12 Operating Manual page 116

Electrocardiograph
Table of Contents

Advertisement

NECG-12
Effective Width:
Paper Speed:
Accuracy of data:
HR Recognition
Technique:
HR Range:
Accuracy:
ECG Unit
Leads:
Acquisition Mode:
A/D:
Resolution:
Time Constant:
Frequency Response:
Gain:
Input Impedance:
Input Circuit Current:
Input Voltage Range
Calibration Voltage:
DC Offset Voltage:
Minimum Amplitude:
Noise:
Multichannel crosstalk
Filter
Electrocardiograph User Manual
210mm
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s
(±3%)
±5% (x-axis), ±5%(y-axis)
Peak-peak detection
30 BPM ~300 BPM
±1 BPM
12 standard leads
simultaneously 12 leads
24 bits
2.52μV/LSB
≥3.2s
0.01Hz ~ 300Hz (-3dB)
1.25mm/mV, 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV,
10/5mm/mV, AGC
≥100MΩ (10Hz)
≤0.01μA
≤±5 mVpp
1mV±2%
±600mV
20 μVp-p
≤12.5 μVp-p
≤0.5mm
AC Filter: On/Off
- 110 -