Chapter 10: Hint Information - EDAN Norav NECG-12 Operating Manual

Electrocardiograph
Table of Contents

Advertisement

NECG-12

Chapter 10: Hint Information

Hint information and the corresponding causes provided by the electrocardiograph are listed in
Table 10-1.
Hint Information
Lead off
Battery Weak
No Paper
Testing
Paper Error
Testing
Sampling/Analyzing/
Recording
Learning
Detecting
Transmitting
Loading Order...
Memory Full
Module Error
DEMO
Overload
U Disk / USB Printer
/ USB Scanner
Electrocardiograph User Manual
Table 10-1 Hint Information and Causes for Resting ECG
Causes
Electrodes fall off the patient or the patient cable falls off the unit ,
or a high polarization voltage occurs.
The battery is weak.
Recorder paper runs out or is not loaded.
The ECG data is being sampled periodically.
When Paper Marker is set to Yes, the electrocardiograph
advances the recorder paper to the next black marker. If it
advances the paper for 300mm and cannot find the next black
marker, the hint Paper Error is displayed.
The ECG data is being sampled periodically.
ECG signals are being sampled / analyzed / recorded.
The self-study process of arrhythmia arithmetic in the Trigger
Sample mode
The examining process of arrhythmia data in the Trigger Sample
mode
ECG data is being transmitted from the electrocardiograph to the
PC through the net or serial cable in the auto or rhythm mode.
Orders are being loaded to the electrocardiograph.
There is no space for saving more records.
There is something wrong with the signal sample module.
The system is in the demonstration mode.
The direct current offset voltage on an electrode is too high.
A U disk, a USB printer or a bar code reader is connected to the
USB interface.
- 95 -