3.3.2 Attaching the Reusable Electrodes (for Resting ECG)

3.3.2.1 Attaching the Limb Electrodes

Connecting to a Lead Wire
Electrocardiograph User Manual
Electrode Placement
V1
Right mid-axillary line on the same horizontal level as C3 and C4
V2
Sternum at the level of C3 and C4
V3
Mid-clavicular line in the fifth intercostals space
V4
Left mid-axilary line on the same horizontal level as C3
V5
Center of spine on the same horizontal level as C3 and C4
V6
Neck, avoid carotid artery and jugular vein
LA
Right arm/Right deltoid
RA
Left arm/Left deltoid
LL
Right leg/Upper leg as close to torso as possible
RL
Left leg/Upper leg as close to torso as possible
Clamp
Limb Electrode Connection:
1) Ensure that the electrodes are clean;
2) Clean the electrode area which is a short distance
3) Daub the electrode area on the limb with gel evenly;
4) Place a small amount of gel on the metal part of the
5) Connect the electrode to the limb, and make sure that
6)
Limb Electrode
above the ankle or the wrist with 75% alcohol;
limb electrode clamp;
the metal part is placed on the electrode area above
the ankle or the wrist;
Attach all limb electrodes in the same way.