Rhythm Mode - EDAN Norav NECG-12 Operating Manual

Electrocardiograph
Table of Contents

Advertisement

NECG-12

6.1.3 Rhythm Mode

The rhythm mode contributes to discover the abiogenetic or frequent arrhythmia. The
rhythm-lead ECG waveform of one lead or three leads can be sampled or printed for a long time.
Operation Method:
1. Set Mode Options, Rhythm Style, Rhythm Lead1/2/3 or Lead Sequence(Optional)
1) Select RHYT from the Mode Option list in the Work Mode Setup window.
2) Select a style from the Rhythm Style list in the Work Mode Setup window.
3) Select a lead from the Rhythm Lead1/2/3 list in the Lead Setup window.
4) Select a sequence from the Lead Sequence list in the Lead Setup window.
5) Press Enter to confirm.
2. Press the PRINT/STOP key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 60s in the Single Lead style or 20s in the
Three Leads style, it begins to print an ECG report.
Electrocardiograph User Manual
- 51 -