Preparation methods according to ISO 17664
Autoclave using the gravitation method for at least 10 minutes at
134°C ± 1°C (273°F ± 1.8°F)
Drying time: 30 min.
Autoclave using the gravitation method for at least 60 minutes at 121
°C ± 1°C (250 °F ± 1.8°F)
Drying time: 30 min.
Follow the manufacturer's instructions for use.
6.10 Storage
Prepared products should be stored protected germ-free from dust in a dry,
dark and cool room.
Note
Observe the expiration date of the sterilized item.
55