Disinfection: Manual Disinfection - Internal - KaVo INTRA 3610 N3 Instructions For Use Manual

Surgical handpiece
Table of Contents

Advertisement

Preparation methods according to ISO 17664
Required tools:
▪ Cloths for wiping down the medical device.
Spray the disinfectant on a cloth, then thoroughly wipe down the med‐
ical device and leave the disinfectant to soak in according to the in‐
structions from the disinfectant manufacturer. disinfectant
manufacturer. .
Note
Observe the instruction for use for the disinfectant.

6.5.2 Disinfection: Manual disinfection - internal

For effective preparation, the interior must be cleaned mechanically using
the cleaning and disinfectant device in agreement with ISO 15883-1.
(The inside of this product should not be disinfected manually.)
43

Advertisement

Table of Contents
loading

This manual is also suitable for:

0.524.5620

Table of Contents