 Acknowledge selection by ➔-button. < Zero >,
< Put ld > followed by the weight value of the
adjustment weight to be placed will be displayed.
 Place the adjustment weight and confirm with ➔-button,
< wait > followed by < reMvld > will be displayed.
 Once < reMvld > is displayed, remove the adjustment
weight.
 After successful adjustment the balance automatically
returns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show the error message < wrong
>. Switch off balance and repeat the adjustment process.