Device-Specific Technical Data - Keller CellaTemp PX1 Series Manual

Pyrometer
Hide thumbs Also See for CellaTemp PX1 Series:
Table of Contents

Advertisement

Operating instructions CellaTemp PX 1x 2x 3x

21 Device-specific technical data

PX 10
Temperature range
Sensor
Spectral sensitivity
Focussing range
Distance to target-size ration
Resolution current output
Resolution temp� reading
Response time t
Measurement uncertainty
Repeatability
Temperature coefficient with
reference to 23°C
0 - 1000 °C
Thermopile
8 -14 µm
0.3 m - ∞ optic PZ 10.01
0.15 - 0.3 m optic PZ 10.05
50:1 (PZ 10.01)
48:1 (PZ 10.05)
0�2 K + 0�03 % of selected range
0,1 K < 200 °C
1 K ≥ 200 °C
≤ 30 ms
98
1 % of temperature reading, minimum 2 K,
(at ɛ =1.0 and Tu = 23 °C)
1 K
≤ 0.1 K / K (for T < 250 °C)
≤ 0.04 %/K (for T ≥ 250 °C)
of temperature reading / K