Summary of Contents for Keller PYRO CellaTemp PA

  • Page 1 PYROMETER CellaTemp ® Mixed-Mode Function Ident no.: 1117971 06/2022...
  • Page 2 Operating manual...
  • Page 3: Table Of Contents

    Operating manual Contents General description................1 Mixed-mode function for extending the measuring range ....1 Configuration layer Mixed-Mode ............2 Factory setting of the mixed mode function ........3 Adjustment of the measured values ..........3 Setting parameters at the pyrometer ..........4 Setting the parameters using the button ............
  • Page 4 Operating manual...
  • Page 5: General Description

    Operating manual General description. The mixed-mode function is used to link two measured values of a ratio pyrometer and to generate a continuous analogue output signal. In the lower measuring range up to the start temperature of the transition range, the measured value of the set source 1 is output. Above the defined end temperature of the transition range, the output signal corresponds to the measured value of source 2.
  • Page 6: Configuration Layer Mixed-Mode

    Operating manual Configuration layer Mixed-Mode The following table the parameters of the mixed-mode function are listed and describes their function. C004 Configuration layer (Mixed-Mode parameter) Parameter Function Explanation MIX. Mixed-mode define lower limit of RNG._ temp. span define upper limit of RNG.~ temp.
  • Page 7: Factory Setting Of The Mixed Mode Function

    Operating manual Factory setting of the mixed mode function When delivered, the pyrometer is already set to mixed-mode operation and the parameters are configured according to the table. Configuration layer C004 Parameter PA 40 AF 1 PA 40 AF 4 PA 50 AF 1 MIX.
  • Page 8: Setting Parameters At The Pyrometer

    Operating manual Setting parameters at the pyrometer Setting the parameters using the button Use the buttons ▲▼ and the „MODE" button on the rear panel to ac- cess and configure parameters. With these buttons you can view and adjust all settings required for pyrometer operation. Menu structure: MODE MODE...
  • Page 9: Main Menue

    Operating manual Enter press Main Menu Execute the [RETURN] command or [ESC] Select [KEY] Backspace Enter press Submenu 1 Execute the [RETURN] command or [ESC] [ESC] Select [KEY] Backspace Press [] Enter sste[] press Submenu 2 Execute the [RETURN] command or [ESC] [ESC] To set the pyrometer to the terminal mode, simultaneously hold down...
  • Page 10: Description Of The Mixed Mode Submenu

    Operating manual Do 1 source ..ready-signal Do 1 function level/signal Unit ....Celsius Do 1 delay time ..0.00 s Terminal assigned to ..USB Do 1 hold time ..0.00 s Autoprint ....on Do 2 source ....off Print cycle time ..
  • Page 11 Operating manual With command „T“ in the main menu you can select the source for the analogue output. Select “4” for Mixed Mode. The display shows the Mixed Mode temperature. >T Set Analog Out 1 SOURCE: 1: Lambda 1 2: Lambda 2 3: Quotient 4: Mixed ------------------------------------------------------------------...
  • Page 12 Copyright prohibits the reproduction or distribution of this instruction manual, in- cluding text, photographs or images contained herein, in whole or in part, for any purpose whatsoever, without prior consent of the author. This applies to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through any medium.