RADIANQBIO INSTRUMENT
2) Compensation functions and notice
If the actual load (load, displacement, pressure) is applied to the sensor and there is a
difference between the value displayed in the indicator and the actual load value, the display
value of the indicator can be changed to the actual load value. Up to 9 points can be stored. As
the calibration interval is differentiated, the linearity of the sensor can be increased. In the actual
load calibration mode, calibration is possible at the forward (+) output of the sensor and
calibration is not possible at the reverse (-) output of the sensor.
If an incorrect value is entered during setup, re-calibration with 8-4 Digital calibration,
delete all data of the values changed to the actual load calibration setting.
Then re-run the actual load calibration setting.
capacity
kg
< Before Compensation >
① Sensor output(display data)
② Real load data
①
②
output
mV/V
35
①
capacity
kg
< After Compensation >
① Sensor output(display data)
② Real load data
③ Values stored with calibration
settings
www.radian21.com
③
②
③
output
mV/V