Radian RI-10W User Manual page 29

Digital indicator
Table of Contents

Advertisement

RADIANQBIO INSTRUMENT
F-32. DAC speed (Set Analog output speed)
Display data
0
Display DAC mode :
High speed DAC mode : Analog output by converting A/D input values
1
(200Hz) (RI-10W:200Hz, RI-50W:1kHz)
NOTE 1) When set to High speed DAC, the analog output has the same linearity as the
calibration setting of 8-3.Digital calibration.8-4.The value set by Actual Load Calibration is not
affected.
NOTE 2) When set to high speed DAC, the analog output is flexible according to the A/D input
value (output of the sensor) and is not affected by the Hold action, F-13 Base offset, or the F-
50 Display reverse mode setting.
NOTE) RI-10W, RI-20W, RI-25W, RI-27W, RI-148A,B
RI-50W
F-33. Zero adjustment (analog output zero adjustment)
Display data
-999
-0.33V(-0.36mA)
999
+0.33V(+0.36mA)
F-34. Span adjustment (analog output range adjustment)
Display data
-999
-1V(-1.6mA)
999
+1V(+1.6mA)
F-40. ID Number (Communication Device Number setup)
Display data
000
000
255
001 ~255
Analog output by converting the current display value
:
key reduction,
When entering the setup, the analog output displays
ZERO value,
and adjusts to
:
key reduction ,
When entering the setup, the analog output displays
SPAN value,
and adjusts to
: Device number is not set-up(Stream mode:always transmit data )
: Device number is set-up(Command mode:Transmit data by command)
28
(Standard setup value : 0)
Setting
: 200Hz
: 1kHz
(Standard setup value : 000)
Setting
key increase
key and
(Standard setup value : 000)
Setting
key increase
key and
(Standard setup value : 000)
Setting
www.radian21.com
key .
key

Advertisement

Table of Contents
loading

Related Products for Radian RI-10W

Table of Contents