RADIANQBIO INSTRUMENT www.radian21.com 1. Features We thank you for using our product. Please refer to this manual or contact our office if you find any problems during using our product. This product is an indicator that emplifies the micro voltage signals from each sensor and display in digital.
RADIANQBIO INSTRUMENT www.radian21.com 2. Attention For the efficient and safe use, please carefully read and be fully aware of the following details before using this product. It is strictly forbidden to use this product for any other purpose of use or to attempt to make any alteration on this product.
RADIANQBIO INSTRUMENT www.radian21.com 4. Front panel ⑴ RI-10W, RI-20W, RI-25W, RI-27W, RI-30W, RI-50W ① comparison output Indication LED: The LED lights up when the Relay is in contact output state. ② HOLD Indication LED: This LED will be lighted when the measured value is on Hold.
Page 8
RADIANQBIO INSTRUMENT www.radian21.com ⑵ RI-148A, RI-148B ① HI, LO Indication LED: LED will be lighted when measured value exceeds the setting value. ⑧ HOLD Indication LED: This LED will be lighted when the measured value is on Hold. ⑨ MOD : Measured value stable (on), unstable (off) indication ④...
5. Rear panel ⑴ RI-10W, RI-20W, RI-25W, RI-27W, RI-30W, RI-50W Please check the location of terminal and its use. While the projected button is being pressed, please insert the cable into the lower hole completely. As soon as you release the button, the connection will be completed.
RADIANQBIO INSTRUMENT www.radian21.com 7. Components & Function 7-1. How to use Hold mode Hold mode is largely divided into Peak Hold and Sample Hold. Please select Peak or Sample. To input Hold, you can use Hold key on the front panel or external input. For the operation method, please refer to the drawing below.
RADIANQBIO INSTRUMENT www.radian21.com 7-2. How to use comparison ouput function For comparison output function, there are 4 different modes such as Decision, High limit, Low limit, Low&High limit. It displays through the relay of rear panel comparing each setup value. On limit mode, Hysteresis can be used.
Page 18
RADIANQBIO INSTRUMENT www.radian21.com * Note) In Function mode, if the value of F-13 is set the Base Offset (page26), the comparative output operates as follows. 1) Decision mode : Measured Value ≤ (Offset-Lower limit setup value) ⇒ RY1 ON (Low) Measured Value ≥...
RADIANQBIO INSTRUMENT www.radian21.com 8. Setting Modes 8-1. Types of Setting mode & Set-up For setting mode, there are 3 different types of mode such as Function mode, Digital calibration mode and Actual load calibration mode. Mesuring mode While key is being Press, press Function mode ①...
RADIANQBIO INSTRUMENT www.radian21.com 8-2. Function mode 1) How to set function (for RI-10W, RI-20W, RI-25W, RI-27W, RI-50W, RI-148A,B) The setting value of Function mode (page 22~29) can be changed the key operation below. (F-00~F-52) Measuring mode While key is being...
RADIANQBIO INSTRUMENT www.radian21.com 2) How to set function (for RI-30W) The setting value of Function mode (page 22~29) can be changed the key operation below. (F-00~F-52) Measuring mode While key is being Press, press Change to Select and save by key.
RADIANQBIO INSTRUMENT www.radian21.com 4) How to set RELAY data ① Decision mode (F-20 Decision mode: mode: 0) In the measuring mode, press the front key(RY1) or key(RY3) to display the currently set value, change it to the desired value, and press the key to save it.
RADIANQBIO INSTRUMENT www.radian21.com 8-3. Function list Name Function Setting Standard setup value Decimal point 0, 1, 2, 3 F-00 Pulse/circle (RI-30W) 0 ~ 360 F-01 Division 1, 2, 5, 10, 20, 50 F-02 Display filter 0.01, 0.02, 0.05, 0.10, 0.20, 0.50, 1.00 0.20 F-03 Hold mode Sample, Peak, Absolute, A/D...
Page 24
RADIANQBIO INSTRUMENT www.radian21.com ※ X : No Function RI-10W RI-20W Name Function RI-30W RI-148 RI-50W RI-25W,27W Decimal point ○ ○ ○ F-00 Pulse/circle (RI-30W) ○ F-01 Division ○ ○ ○ ○ F-02 Display filter ○ ○ ○ ○ F-03 Hold mode ○...
Page 25
RADIANQBIO INSTRUMENT www.radian21.com F-00. Decimal point (Decimal point Set-up) ① Decimal point : RI-10W, RI-20W, RI-25W, RI-27W, RI-50W, RI-148A,B (Standard setup value : 1) Display data 00000 : No decimal point 0000.0 : One decimal place 000.00 : Two decimal places 00.000 : Three decimal places...
Page 26
Note) When set to A/D Peak Hold (+), the Hold operation is performed according to the A/D input (output of the sensor), regardless of the sign of the F-50 Display reverse mode. RI-10W, RI-20W, RI-25W, RI-27W, RI-148A,B : 200Hz RI-50W : 1kHz F-04.
Page 27
RADIANQBIO INSTRUMENT www.radian21.com F-12. Auto zero during operation (Setting Zero when operationg) (Standard setup value : 0) Display data Setting Not used Autozero only once after power-on F-13. Base offset (Setting value) Base (Standard setup value : 00000) Display data Setting -19999 00000...
Page 29
NOTE 2) When set to high speed DAC, the analog output is flexible according to the A/D input value (output of the sensor) and is not affected by the Hold action, F-13 Base offset, or the F- 50 Display reverse mode setting. NOTE) RI-10W, RI-20W, RI-25W, RI-27W, RI-148A,B : 200Hz RI-50W : 1kHz F-33.
Page 31
RADIANQBIO INSTRUMENT www.radian21.com F-51. Force unit (Conversion unit set up) (Standard setup value : 0) Display data Setting kg/㎠ kgf × 9.8 ℓb kgf × 2.2 kg/㎠ × 0.98 kg/㎠ × 0.098 Note1) When setting the conversion unit (set value: 1 to 4), calibration modes of 8-3 and 8-4 are not available.
8-4. Digital calibration (Calibration by sensor output value) 1) RI-10W, RI-27W, RI-50W, RI-148A,B At the time of purchasing sensor, the rated capacity (R.C) and rated output (R.O) declared on the calibration sheet can be used for the calibration for easier calibration.
Page 33
RADIANQBIO INSTRUMENT www.radian21.com 2) RI-20W At the time of purchasing sensor, the rated capacity (displacement) and rated output (R.O) declared on the calibration sheet can be used for the calibration for easier calibration. Refer to 8-5.Actual load calibration if you need to correct the error. Measuring mode While key is being Press, press...
Page 34
RADIANQBIO INSTRUMENT www.radian21.com 3) RI-25W At the time of purchasing sensor, the rated capacity (displacement) and rated output (R.O) declared on the calibration sheet can be used for the calibration for easier calibration. DP-xx series (Half type (HBT) recommends setting the rated output value to approximately 0.3300.
Page 35
RADIANQBIO INSTRUMENT www.radian21.com 8-5. Actual load calibration (Exact Load Calibration) 1) Compensation Calibration (a) Please calibrate by 8-4 Digital Calibration. And, if there is a difference from the actual load value, you can change the indicator's display value to the actual load value by referring to the actual load calibration setup flow chart below.
Page 36
RADIANQBIO INSTRUMENT www.radian21.com 2) Compensation functions and notice If the actual load (load, displacement, pressure) is applied to the sensor and there is a difference between the value displayed in the indicator and the actual load value, the display value of the indicator can be changed to the actual load value. Up to 9 points can be stored. As the calibration interval is differentiated, the linearity of the sensor can be increased.
Page 37
3) If the output of sensor is unknown ① Notes for calibrating RI-10W, RI-27W, RI-50W, RI-148 Calibrate with 8-4 Digital Calibration. Deletes the data of the actual load calibration value. Then enter the output and capacity values as default (3.0000/30000).
Page 38
RADIANQBIO INSTRUMENT www.radian21.com ② Notes for calibrating RI-25W Set calibration with 8-4 Digital Calibration to delete the data of the values changed by the actual load calibration. At this point, the rated output value is 0.6%, and the rated capacity (displacement) of the sensor is set for the sensor.
RADIANQBIO INSTRUMENT www.radian21.com 8-5. DATA BACK-UP & RESTORE You can save all the set-up values of the device and then restore them to the current set-up state as per your need. ◦ DATA BACK-UP : Save the current set-up state. ◦...
RADIANQBIO INSTRUMENT www.radian21.com 8-6. Lock Set-up You can prevent any accidental operation due to the unnecessary key control by Lock set-up. After finishing calibration, it is recommended to set the Lock. At the first stage, please start while the power is OFF. Related Function when Lock is set : Function related to calibration.
Page 42
RADIANQBIO INSTRUMENT www.radian21.com When the measured • Faulty cable connection • Prove proximity sensors operation (Lamp) value does not • Faulty detection of • Check the distance between the proximity increase or proximity sensors sensor and the gear (approximately 1mm) increases slightly •...
RADIANQBIO INSTRUMENT www.radian21.com 10. OPTION 10-1. BCD OUT INTERFACE This Programmable BCD interface is the output for the weight value made into BCD coding. This interface is available for PLC(Programmable Logic Control), Computer. PIN Arrangement PIN No. SIGNAL PIN No. SIGNAL Negative Polarity 1 ...
Page 44
RADIANQBIO INSTRUMENT www.radian21.com Signal output DATA output TIME(T) can set to Function F-04 BCD Mode. (50, 100, 200, 500, 1000㎳) Signal logic ① BCD DATA output : Negative logic (Negative) ② BUSY output : DATA READ = L ③...
RADIANQBIO INSTRUMENT www.radian21.com 10-2. Serial communication Since RS232C Interface is very sensitive of electric noise. So please do the wiring from AC Power and electric wires separately. Also you must use the shield calbe always. 1) Wiring ① Option-02 (RS232C) Indicator Host PC TXD(Transmission Data)
Page 48
RADIANQBIO INSTRUMENT www.radian21.com ① Stream mode (F-40, ID Number setup value ‘000’) (Ex. Data +1234.5 transmieeion) CODE BYTE1 BYTE2 BYTE3 BYTE4 BYTE5 BYTE6 BYTE7 BYTE8 ASCII CODE BYTE9 BYTE10 BYTE11 BYTE12 BYTE13 BYTE14 BYTE15 BYTE16 ASCII 1) BYTE1, BYTE2 : DATA stable (S T) DATA unstable (U S) DATA overflow (O L) DATA underflow (U L)