Radian RI-10W User Manual page 28

Digital indicator
Table of Contents

Advertisement

RADIANQBIO INSTRUMENT
F-30. DAC mode (Setting Analog output type)
① RI-10W, RI-25W, RI-27W, RI-50W, RI-148A,B
Display data
b_05v
-5 ~ +5V output
b_10v
-10 ~ +10V output
b_02A
0 ~ 20mA output
b_42A
4 ~ 20mA output
U_05v
0 ~ 5V output
U_10v
0 ~ 10V output
U_02A
0 ~ 20mA output
U_42A
4 ~ 20mA output
② RI-20W, RI-30W
Display data
b_05v
0 ~ 5V
b_10v
0 ~ 10V
b_02A
0 ~ 20mA
b_42A
4 ~ 20mA
U_05v
0 ~ 5V
U_10v
0 ~ 10V
U_02A
0 ~ 20mA
U_42A
4 ~ 20mA
F-31. DAC capacity (Analog Outupt value setup)
Display data
0
Set the rated capacity of the Analog output
< If Setting value 10000>
Display
+99999
data
-10000
0
10000
Note) Models RI-20W and RI-30W cannot be negative (-) output.
output
output
output
output
output
output
output
output
b_05v
b_10v
b_02A
-5V
-10V
0
0
0
0
+5V
+10V
20mA
27
(Standard setup value : b_10V)
Setting
bidirectional, polarity
unidirectional, unipolar
bidirectional,
(Standard setup value : b_10V)
Setting
unidirectional, unipolar
bidirectional,
(Standard setup value : 30000)
Setting
setting data
F-30
b_42A
U_05v
U_10v
4mA
0
4mA
2.5V
5V
20mA
5V
10V
www.radian21.com
unipolar
unipolar
U_02A
U_42A
0
0
4mA
10mA
12mA
20mA
20mA

Advertisement

Table of Contents