Radian RI-10W User Manual page 26

Digital indicator
Table of Contents

Advertisement

RADIANQBIO INSTRUMENT
F-03. Hold mode
Display data
0
Sample Hold :
1
Peak Hold (+) :
2
Absolute Peak Hold(+/-) :
values during Hold signal input.
3
A/D Peak Hold(+):
Note) When set to A/D Peak Hold (+), the Hold operation is performed according to the A/D input (output
of the sensor), regardless of the sign of the F-50 Display reverse mode.
RI-10W, RI-20W, RI-25W, RI-27W, RI-148A,B
RI-50W
F-04. BCD Busy time (BCD Output time setup)
Display data
0.050
0.100
0.200
0.500
1.000
F-10. Auto zero tracking (Auto zero operation range setup)
Display data
00
00
99
01 ~99
F-11. Auto zero tracking time (Auto Zero tracking time setup)
Display data
0.0
0.0 ~5.0
5.0
To hold the display value at the time of Hold signal input.
To hold the maximum value of display values during Hold signal input.
To hold the maximum absolute value of display
To hold the maximum A/D value during Hold signal input.
50 ms
100 ms
200 ms
500 ms
1000 ms
: Auto zero is not used
: Set up the operation range of auto zero
(If the indicator value is less than the set value, it
corresponds to the autozero operating range.)
: set up the auto zero operation time (0.0 ~ 5.0 sec)
(
When the indicator is in the autozero operating range, the display is
maintained by the operating time setting, and then autozero is
performed.)
Setting
: 200Hz
: 1kHz
(Standard setup value : 0.100)
Setting
(Standard setup value : 00)
Setting
(Standard setup value : 0.0)
Setting
25
www.radian21.com
(Standard setup value: 1)
Hold.