Shodex PROTEIN LW-803 Operation Manual

Advertisement

Quick Links

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)

Important Handling Instructions

Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.

Before Using the Column

(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.

1. Introduction

Thank you for purchasing the Shodex product. Shodex PROTEIN LW-803 is a silica-based gel filtration
chromatography column. The column is packed with particles having carefully optimized pore-sizes. It is suitable
for analyzing proteins with a molecular weight of several hundred of thousand.

2. Column Components

Please refer to the Shodex website: https://www.shodex.com/en/da/07.html

3. Column Specifications

Product
Product Name
Code
PROTEIN LW-803
F6989303
PROTEIN LW-G 6B
F6700133
Base Material
: Spherical porous particles of silica modified with hydrophilic group
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water

4. Usable Conditions

Product Name
PROTEIN LW-803
PROTEIN LW-G 6B
Operation Manual
Shodex PROTEIN LW-803
Column Size
Particle
(mm)
Size
(μm)
I.D.
Length
8.0
300
3
6.0
50
3
Flow Rate (mL/min)
Recommended
Maximum
1.0
-
Resonac Corporation (https://www.shodex.com/)
Theoretical
Target Molecular
Plate Number
Weight Range
(Per Column)
≥ 12,000
10,000 - 700,000
(Guard Column)
Reference value only / Measured with proteins
*1
Maximum
Pressure
(MPa/column)
1.5
15
-
-
ver.20221202E
Exclusion
*1
Limit
*1
(1,000,000)
*2
-
-
*2
Estimated value
pH
Temperature
Range
Range (°C)
3.0 - 7.5
4 - 45
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROTEIN LW-803 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex PROTEIN LW-803

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex PROTEIN LW-803 is a silica-based gel filtration chromatography column. The column is packed with particles having carefully optimized pore-sizes. It is suitable for analyzing proteins with a molecular weight of several hundred of thousand.
  • Page 2 This may lead to abnormal peak shapes, insufficient separations, and/or low repeatability. To avoid the negative influences, it is recommended to lower the sample concentration or alternatively, lower the injection volume. Resonac Corporation (https://www.shodex.com/)
  • Page 3 Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. Resonac Corporation (https://www.shodex.com/)
  • Page 4 · Never allow inside the column to dry. It can damage the column. 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...
  • Page 5 (3) Please follow a proper waste disposal method specified by your local regulations. Please refer to the Shodex website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex support office (https://www.shodex.com/en/support_office/list).

This manual is also suitable for:

F6989303F6700133

Table of Contents