Shodex Asahipak GS-310 20G Operation Manual

Advertisement

Shodex
(Please read this manual carefully to achieve the best and consistent column performance for a long time)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column adhesive label, and
enclosed Certificate of Analysis (CoA) are matching and correct.
1. Introduction
Thank you for purchasing the Shodex™ product. Shodex Asahipak GS-310 20G and GS-510 20G are
polymer-based size exclusion chromatography columns. They work well with both aqueous and organic solvents.
2. Column Components
3. Column Specifications
Product
Product Name
Code
F6810038
Asahipak GS-310 20G
F6810039
Asahipak GS-510 20G
Asahipak GS-10G 7B
F6710020
Base Material
: Spherical porous particles of polyvinyl alcohol
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water/Methanol = 70/30
Operation Manual
Asahipak
GS-310 20G, GS-510 20G
Column Size (mm)
I.D.
Length
20.0
500
20.0
500
7.5
50
SHOWA DENKO K.K. (https://www.shodex.com/)
Particle
Theoretical
Size
Plate Number
(μm)
(Per Column)
13
≥ 14,000
13
≥ 14,000
20
(Guard Column)
*
Reference value only / Measured with pullulan
ver.220208E
Target Molecular
Exclusion
Weight Range
*
Limit
*
300 - 30,000
40,000
5,000 - 200,000
300,000
-
-
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Asahipak GS-310 20G and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex Asahipak GS-310 20G

  • Page 1 Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex™ product. Shodex Asahipak GS-310 20G and GS-510 20G are polymer-based size exclusion chromatography columns. They work well with both aqueous and organic solvents.
  • Page 2 Flow Rate (mL/min) Temperature Product Name pH Range Pressure (MPa) Range (°C) Recommended Maximum (Per Column) Asahipak GS-310 20G 5 - 8 Asahipak GS-510 20G 2 - 9 4 - 60 Asahipak GS-10G 7B 4.2 Usable Solvents Product Name Solvents...
  • Page 3: Eluent Preparation

    · If the eluent left in the system is not compatible with the column to be used, it may damage the column. · A drastic change in the eluent compositions may remove substances adsorbed on the system and they may enter and deteriorate the column. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 4: Column Installation

    When replacing THF to DMF with LiBr, first follow step (3) above, and then replace 100 % DMF to DMF with LiBr. Attention! · Frequent solvent replacement deteriorates the column, and thus not recommended. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 5: Column Storage

    9. Column Inspection Inspection method is described in the Certificate of Analysis (CoA). Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations. (1) Theoretical Plate Number (N) (2) Asymmetry Factor (Fas) SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 6 (2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
  • Page 7 11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.