(Please read this manual carefully to achieve the best and consistent column performance for a long time)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column name tag, and
enclosed Certificate of Analysis (CoA) are matching and correct.

1. Introduction

Thank you for purchasing the Shodex product. Shodex RSpak DS-413 is a polymer-based reversed phase
chromatography column. It can be used in a wide pH range from acidic to alkaline (pH 1 - 13).

2. Column Components

3. Column Specifications

Product
Product Name
Code
F7001012
RSpak DS-413
Base Material
: Spherical porous particles of styrene divinylbenzene copolymer
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water/Acetonitrile/THF = 40/30/30

4. Usable Conditions

Product Name
Recommended
RSpak DS-413
Operation Manual
TM
Shodex
Column Size (mm)
I.D.
4.6
Flow Rate (mL/min)
Maximum
0.6 - 1.0
1.0
*
SHOWA DENKO K.K. (https://www.shodex.com/)
TM
RSpak
DS-413
Particle
Size (μm)
Length
150
Maximum Pressure
(MPa)
(Per Column)
20
Eluent without THF, the maximum usable temperature is 70 °C.
Theoretical Plate Number
(Per Column)
3.5
≥ 11,000
Temperature
pH Range
Range
1 - 13
ver.211129E
(°C)
*
20 - 50
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RSpak DS-413 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex RSpak DS-413

  • Page 1 Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex RSpak DS-413 is a polymer-based reversed phase chromatography column. It can be used in a wide pH range from acidic to alkaline (pH 1 - 13).
  • Page 2 · If the eluent left in the system is not compatible with the column to be used, it may damage the column. · A drastic change in the eluent compositions may remove substances adsorbed on the system and they may enter and deteriorate the column. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 3 (2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate less than 0.3 mL/min. Introduce 5 to 10 times column volume of the washing solvent. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 4 (2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
  • Page 5 11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.

This manual is also suitable for:

F7001012

Table of Contents