(Please read this manual carefully to achieve the best and consistent column performance for a long time)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column name tag, and
enclosed Certificate of Analysis (CoA) are matching and correct.
1. Introduction
Thank you for purchasing the Shodex™ product. Shodex PROTEIN KW-800 series is a silica-based gel
filtration chromatography column. The column series is suitable for the size exclusion separations of proteins.
The columns with different exclusion limits are available. Please select columns with target molecular weight
ranges (see below) that meet the molecular weight range of your samples.
2. Column Components
3. Column Specifications
Product
Product Name
Code
F6989000
PROTEIN KW-802.5
F6989103
PROTEIN KW-803
F6989104
PROTEIN KW-804
F6700131
PROTEIN KW-G 6B
Base Material
: Spherical porous particles of silica modified with hydrophilic group
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water
Operation Manual
Shodex
PROTEIN KW-800 series
Column Size
Particle
(mm)
Size (μm)
I.D.
Length
8.0
300
5
8.0
300
5
8.0
300
7
6.0
50
7
SHOWA DENKO K.K. (https://www.shodex.com/)
Theoretical
Target Molecular
Plate Number
Weight Range
(Per Column)
≥ 21,000
5,000 - 100,000
≥ 21,000
10,000 - 700,000
≥ 16,000
30,000 - (4,000,000)
(Guard Column)
*1
Reference value only / Measured with protein
ver.211111E
Exclusion Limit
*1
150,000
(1,000,000)
*2
*2
(4,000,000)
*2
-
-
*2
Estimated value
*1
1