(Please read this manual carefully to achieve the best and consistent column performance for a long time)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column name tag, and
enclosed Certificate of Analysis (CoA) are matching and correct.

1. Introduction

Thank you for purchasing the Shodex product. Shodex IC YS-50 is an ion chromatography column
suitable for a simultaneous analysis of monovalent and divalent cations. It is applicable for both suppressor
and non-suppressor methods.

2. Column Components

3. Column Specifications

Product
Product Name
Code
F7122000
IC YS-50
F6700530
IC YS-G
Base Material
: Spherical porous particles of polyvinyl alcohol modified with carboxyl group
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : 10 mM Sodium Sulfate aq. Solution
Operation Manual
Shodex
Column Size (mm)
I.D.
4.6
4.6
SHOWA DENKO K.K. (https://www.shodex.com/)
IC YS-50
Particle
Size (μm)
Length
125
5
10
5
ver.211201E
Theoretical Plate Number
(Per Column)
≥ 5,500
(Guard Column)
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IC YS-50 and is the answer not in the manual?

Questions and answers

Summary of Contents for Shodex IC YS-50

  • Page 1 Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex IC YS-50 is an ion chromatography column suitable for a simultaneous analysis of monovalent and divalent cations. It is applicable for both suppressor and non-suppressor methods.
  • Page 2 · If the eluent left in the system is not compatible with the column to be used, it may damage the column. · A drastic change in the eluent compositions may remove substances adsorbed on the system and they may enter and deteriorate the column. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 3 (2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at 0.5 mL/min. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 4 9. Column Inspection Inspection method is described in the Certificate of Analysis (CoA). Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations. (1) Theoretical Plate Number (N) (2) Asymmetry Factor (Fas) SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 5 (2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
  • Page 6 11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.