9

Meter Function Menus

MENU
BUTTON
MAIN MENU
"L1 CNF":
*The 'MIN' button allows you to sequence through L1C.1, L1C.2, L1C.3,
L1C.4, L1C.5, L1C.6, L1C.7 and L1C.8.
The 'MAX' button allows you to select the "0" or "1" state for each "L1C"
condition.
The 'MENU' button stores the selected values for all "L1C" condition(s)
changed and advances the meter to "L2 CNF". Do not press the 'MENU'
button after each change within the submenu or the meter will advance to
the next menu item.
Every underlined "0" or "1" state is the factory preset value.
9-2
MIN/MAX*
BUTTON
SUB MENU
CONDITION
LOCKOUT CONFIGURATION #1
"L1C.1=0"
Setpoint 1 change unlocked.
"L1C.1=1"
Setpoint 1 change locked out.
"L1C.2=0"
Setpoint 2 change unlocked.
"L1C.2=1"
Setpoint 2 change locked out.
"L1C.3=0"
Setpoint 3 (Alarm 1) change unlocked.
"L1C.3=1"
Setpoint 3 (Alarm 1) change locked out.
"L1C.4=0"
Setpoint #4 (Alarm 2) change unlocked.
"L1C.4=1"
Setpoint #4 (Alarm 2) change locked out.
"L1C.5=0"
Valley-value (LO RDG) display is permitted.
"L1C.5=1"
Displays microprocessor revision.
"L1C.6=0"
Peak-value (HI RDG) display is permitted.
"L1C.6=1"
Peak-value (HI RDG) display is not
permitted.
"L1C.7=0"
INPUT CLASS (BRIDGE) is selected.
"L1C.7=1"
INPUT CLASS is locked out.
"L1C.8=0"
Not used in "BRIDGE" mode,
must be set to "0".
"L1C.8=1"
Not used in "BRIDGE" mode.