Thermo Scientific 1500 Series Operating Instructions Manual page 44

Biological safety cabinet
Table of Contents

Advertisement

Table 12. QuickChart – 1500 Series Type B2
Model (Series)
Exhaust HEPA Leak Test Data
Air Displacement (CFM)
Laskin Nozzles needed
Theoretical aerosol conc. (ug/l)
Actual aerosol conc. (ug/l)
Exhaust Data
DIM Exhaust Volume (CFM)
Traverse Exhaust Volume (CFM)
Differential Pressure (in. H
Exhaust (Inflow) Alarm Data
Alarm Setpoint (CFM Inflow)
Supply HEPA Data
Thermo P/N
Width x Depth x Height (in.)
Performance (CFM)
Performance (Pressure in. H
Exhaust HEPA Data
Thermo P/N
Width x Depth x Height (in.)
Performance (CFM)
Performance (Pressure in. H
Motor/Blower Data
7
Thermo P/N
Motor HP
LED Lamp Data (2 each)
Thermo P/N
Color (ºK)
Lumens
Glass Type
1500 Series Class II, Type B2 Biological Safety Cabinet
1
2
3
4
5
0)
2
6
O)
2
6
O)
2
8
Part Numbers
1510, 1511, 1515, 
1516, 1517, 1518
723
2
38
21
723
852
1.8
210 +/- 5
3838401
48" x 18" x 3.06"
445
.5+/-.1
3438501
26" x 18" x 8.08"
720
.48+/-.07
3331104
½
1297505
4000
1850
Frosted
1560, 1561, 1562,
1563, 1564, 1564
1083
2
25
17
1083
1265
2.2
315 +/- 5
3838403
72" x 18" x 3.06"
665
.38-.5
3438503
48" x 18" x 8.08"
1100
.37+/-.07
3331105
½
1297506
4000
3200
Frosted
Appendix E: Quick Chart |
39