CLOSE SASH:
UV LIGHT ON
Figure 31. Cabinet UV light setting
Whether the UV Light is initiated from the [UV Light] button
on the keypad, or automatically initiated upon closing of the
sash, this screen controls the time the UV lamp will remain on.
Select the length of time desired, press [OK/Mute].
UV OPERATION
(MINUTES)
Figure 32. UV Operation
22 |
Using the Cabinet
Navigating the Settings
Submenu
This submenu allows you to select: Units of Measure, System
Lock, Data Output, or UV Parameters.
Units of 
Measure
System Lock
Data Output
Figure 33. Submenu settings
Selecting the Units of Measure
If your cabinet is equipped with an airflow sensor, the units of
measure can be set for FT/MIN (feet per minute) or M/S
(meters per second). Select the appropriate units of measure,
then [OK/Mute].
METRIC (M/S)
IMPERIAL (FT/MIN)
Figure 34. Units of measure
Activating the Security Lock
The Security Lock "locks" the keypad to prevent unauthorized
use of the cabinet. To enable / disable select Protected /
Unprotected, then [OK/MUTE]. When enabled, the keypad is
locked immediately after the blower is turned off. The security
lock is deactivated by holding [DOWN] for three seconds. If
blower is not turned on within 5 minutes of unlocking, the
1500 Series Class II, Type B2 Biological Safety Cabinet
UV Parameters