Remote Operation (Computer)
SBI functions for remote operation
are useful if the current device/application
status can be read out at any time. This
is achieved using the ESC ars_ command
(see next page). The response from the
device to the ESC ars_ command has the
format: XXX xx zz ff.
Device response Meaning
XXX
Active application
xx
Application status
zz
Cover status
ff
F unction of the u
key
xx
Operating status
1
TARE
2
WEIGHING
3
WEIGHING
4
TARE/MEASUREMENT START Waiting for stability
5
TARE
6
START MEASUREMENT
7
MEASUREMENT
8
MEASUREMENT PAUSE
9
MEASUREMENT END
14
SETUP/PROGRAM
15
ERROR
zz
Meaning
O
Cover open
C
Cover closed
XXX
MAN
CBA
CHT
CMS
THW
Action
Tare sample pan
Apply sample
Start analysis
Waiting for tare to complete Dark
Close sample chamber,
delay time running
Analysis active
Analysis cover open
Analysis ended
Configure parameters
L, H, scale error 50, 53, 54
ff
Meaning
0
No soft key
1
Soft key 1
2
Soft key 2
3
Soft key 3
Meaning
Moisture analysis/Setup
menu/Program menu
Cal./adj. scale
Heater adjustment
M easurement
adjustment
Hardware test
Main display
Weight value
Weight value
Weight value
Dark
Weight value
Result
Weight value
Result
Menu display
L, H, ERR xxx