Step
Printout following adjustment:
13. Unload the weighing system
34
Key (or instruction)
Display/Printout
--------------------
14.07.2006 15:15
Model MA150Q-000230V
Ser. no. 18701636
Vers. no. 01-50-01
ID
--------------------
External calibration
W-ID
Nom. + 100.000 g
Diff. - 0.002 g
External adjustment
completed
Diff. 0.000 g
--------------------
14.07.2006 15:16
Name:
--------------------