Display Mode
The following units can be selected for
displaying analysis results:
– Moisture
%L
– Dry weight
%R
– Ratio
%LR
– Weight loss
mg
– Residual weight g
– Residual weight g/kg
– Gram/liter
g/l
Print Intermediate Results
– When set to "Off", intermediate
results can only be printed by pressing
the r key.
– Intermediate results can be printed out
automatically after a configurable time
interval.
Determination of Ignition Residue
A two-part measurement can be
performed using this setting. Example:
Moisture analysis and incineration
of sludge. Moisture content is analyzed
in the first stage. The sample is then
reduced to ashes in the furnace and
in the second stage the ash is calculated
as a percentage of the initial weight.
is displayed instead of
CONTINUE
after the first stage.
The moisture analyzer must not be used
for other purposes between the first and
second stages.
Headers
Text can be entered for the drying
report (2 lines each with max.
20 characters). The relevant line is
not printed if no text is entered.
Factory Setting
The drying programs can be reset to
the factory settings.
26
Printout of the Analysis Parameters
§ When the desired program is displayed:
Press the r key
> Printout (example on right). Lines with
more than 20 characters are truncated.
Additional Functions
The following functions are supported
in the program memory:
– Display programs
– Change settings
– Save program changes
END
--------------------
14.07.2006 12:40
Model MA150Q-000230V
Ser. no. 18701636
Vers. no. 01-50-01
ID
--------------------
PROGRAM
--------------------
P1
Program name
Progr. name:
MEYER
Heating program
Standard drying
Temperature:
105 °C
Standby temperature
Off
Initial weight
Off
Start of measuremen
Without stability,
after close
End of measurement
Fully automatic
Result display
Moisture (%L)
Print intermediate
res.
Off
Determination of
ignition residue
Off
Headers
Line 1
Line 1:
Line 2
Line 2:
--------------------