Sartorius MA150 Operating Instructions Manual page 33

Electronic moisture analyzer
Hide thumbs Also See for MA150:
Table of Contents

Advertisement

Step
Weighing system is being prepared
for calibration
The following is displayed when the system
is ready for calibration
10. Place the 100 g calibration weight
on the weighing system (here 100.00 g)
"–" sign: weight too low
"+" sign: weight too high
No +/- sign: weight ok
At the end of calibration the display shows
11. If no adjustment is required,
select
in the Function line
END
and confirm
   
   
   
   
   
   
   
   
   
   
   
   
   
   
12. If necessary, adjust the weighing system
At the end of adjustment, the display shows
a nd a report is printed: (see next page)
Key (or instruction)
Position the 100 g
calibration weight
y
u
u
Display/Printout
WGH.SYS
CAL./ADJ.Verzög. 2sec
g
100000
D
SETUP
PROGRAMM
WGH.SYS
CAL./ADJ.Verzög. 2sec
g
100000
D
SETUP
PROGRAMM
WGH.SYS
CAL./ADJ.Verzög. 2sec
0002
g
D
SETUP
END
--------------------
14.07.2006     15:10 
Model MA150Q-000230V 
Ser. no.    18701636 
Vers. no.   01-50-01 
ID 
-------------------- 
External calibration 
W-ID 
Nom.  +  100.000 g 
Diff. -     0.002 g 
-------------------- 
14.07.2006          15:12 
Name: 
--------------------
WGH.SYS
CAL./ADJ.Verzög. 2sec
0002
g
D
SETUP
ENDE
WGH.SYS
CAL./ADJ.Verzög. 2sec
g
100000
S
END
CAL.
TARA
TARA
ADJUST
JUST.
TARE
33