Replacing The Valve Plates - M&C TEC1 Series Operation Manual

Thermoelectric chiller
Table of Contents

Advertisement

15.1

Replacing the valve plates

Note
For replacing the valve plates, the sample gas fitting
To replace the valve plates, proceed as follows:
1. Unscrew the crankcase cover. To do this, loosen the 3
hexagon socket screws F (3 mm spanner).
2. Loosen the four hex socket screws G (3 mm spanner).
3. Remove pressure ring H.
4. Remove upper pump head A.
Caution
Turn pump head A only to the right and up-
wards. This prevents the bellows from loosen-
ing from the connecting rod.
O-rings C and valve plates B are freely accessible now.
5. The valve plates and O-rings can now be cleaned or re-
placed.
6. Clean valve seats and pump head with an adequate sol-
vent (e.g. alcohol) and use oil-free compressed air to re-
move dust particles from the parts.
After replacing or cleaning the valve plates and O-rings, follow these steps to re-assemble the pump:
1. Insert O-rings C into the grooves, and place valve plates B back into the cleaned valve seats. Make
sure that the valve plates are in the correct position.
Note
www.mc-techgroup.com
If you replace the valve plates, we also recommend to replace the O-
rings C at the same time.
On the pressure side, the smooth side of Valve B is facing downwards,
and on the suction side the smooth side of Valve B is facing upwards
(operating direction is marked with an arrow on the pump head).
MP-
A
B
C
F
Figure 9
Sectional drawing MP-F..
| 1.04.01
.
D
E
H
G
25