Wrong Operation - M&C TEC1 Series Operation Manual

Thermoelectric chiller
Table of Contents

Advertisement

Wrong operation

Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
8
Please make sure to install and operate the pump for the in-
tended use described in this instruction manual only.
The pump can only be operated within the parameters stated in
The pump is not designed for liquids or particles.
It is prohibited to operate the pump with condensate.
The pump head can be electrostatically charged by particles,
liquids and drops of condensate.
Condensate accumulated inside the pump head destroys the
pump.
If it is likely that humidity or condensation occurs, the pump
head needs to be rotated 180° facing downwards. With the
pump head facing downwards, the condensate can leave the
pump head.
The materials of the pump, which are in contact with the sample
gas, need to be suitable for the used sample gas.
In order to operate the pump, it needs to be securely installed.
Install the pump only in protected areas, sheltered from rain
and moisture.
Please make sure to provide sufficient ventilation.
Do not exeed the permitted maximum pressure within the pump
(gauge pressure).
Do not close the sample gas output. A sufficient sample gas
flow needs to be provided to prevent the pump pressure from
exceeding the permitted maximum pressure value.
A sufficient sample gas flow also prevents the temperature of
the pump head to increase over the permitted temperature
range.
If components downstream from the pump can block or reduce
the sample gas flow, appropriate measure, e.g. a pressure
valve, needs to be installed to prevent the process pressure
from exceeding the maximum permitted pressure value.
MP-
4.01