Definitions for all parameters are listed as below:
Parameter
V Limit
I Limit
OCP Delay
Voltage Level
Voltage Sense
Degree On
Degree Off
Is Start
Is Interval
Waveform
Buzzer
Keylock

3.5.1 Voltage Limit

To limit the maximum output voltage for all pages, Vdc Limit P and Vdc Limit N limit the
maximum positive and negative DC voltage. It is related to Voltage Level parameter,
resolution is 0.1V:
It is Firmware to set the limitation instead of by Hardware.
20
│ www.deltaww.com
back to previous page
Page down
Definition
Maximum voltage setting for all
pages
Maximum output Irms
Delay time for over current protection
Output voltage range
Output voltage measurement point:
AC source output/ input of testing
product
Start phase angle of output voltage
Stop phase angle of output voltage
Delay time to start measuring output
surge current
Duration to measure surge current
Output waveform set at main manual
Toggle switch for Buzzer
Lock settings at main manual
Setting Range
Min
Max
0.0
150.0 / 300.0
0.0
16.0 / 8.0
0.0
5.0
150 V / 300 V
Volage Sense /
Remote Sense
0.0
359.9
0.0
360.0
0.0
1000.0
0.0
1000.0
A / B
ON / OFF
ON / OFF
Unit
V
A
Sec
degree
degree
msec
msec