Air travel requirements
This device is not suitable for use in the presence of a flammable anesthetic mixture.
Notes:
The manufacturer reserves the right to change these specifications without notice.
Pressure may be displayed in cm H
Displayed values
Value
Flow sensor in device
2,3
Leak
Tidal volume
Respiratory rate
Minute ventilation /
Alveolar minute
2
ventilation
Ti
Pressure
1 Results may be inaccurate in the presence of leaks and may be affected by vent characteristics.
2 Results are expressed at STPH (Standard Temperature Pressure, ambient humidity)
3 Results may be inaccurate in the presence of leak and supplemental oxygen > 10 L/min. If necessary for
therapy, use external monitoring.
4 Results may be inaccurate if the tidal volume is below 50 mL.
Pressure-flow table
Medical-Portable Electronic Devices (M-PED) that meet the Federal
Aviation Administration (FAA) requirements of RTCA/DO-160 can be
used during all phases of air travel without further testing or
approval by the airline operator. ResMed confirms that the Stellar
150 meets RTCA/DO-160 requirements.
O or hPa.
2
Range
1
0–120 L/min
2,3
50–3,000 mL
5–60 bpm
0.6–60 L
0.0–4 sec
0-40 cm H
O
2
Stellar 150 with 2m standard tube
Flow (L/min)
Accuracy
± 12 L/min or ± 20%,
whichever is greater
± 20% or 30 mL whichever
is greater
4
±1 bpm
± 0.5 L or 20% whichever is
greater
± 0.1 sec or ±10%,
whichever is greater
IPAP: ± 0.5 cm H
O ± 10%
2
of set pressure (end of
inspiration);
EPAP or PEEP: ± 0.5 cm
H
O ± 4% of set pressure
2
CPAP: ± 0.5 cm H
O ± 10%
2
of set pressure
Technical specifications
Display Resolution
1 L/min
10 mL
1 bpm
0.1 L
0.1 sec
0.5 sec
65