Parameter
Flow Unit
Therapy LED
Confirm Stop
Pressure
Naming
Multi Program
Height Unit
Erase Data
To return to the Options screen press the Return button
1 Parameter appears in both the patient and clinical mode.
44
Default
Description
L/min
Selection of the display units for flow.
Options: L/min, L/sec
Off
Enables or disables the therapy LED. If enabled, the blue therapy LED
in the Start/Stop button is lit while the device is delivering therapy.
Options: On, Off
On
This feature helps to avoid anyone inadvertently stopping therapy.
When enabled (On), if you press
screen will appear. If YES is selected, therapy stops. If NO is selected
or any other key is pressed, therapy continues.
Options: On, Off
IPAP/
Choose the terminology type used through the screens displays.
EPAP
Options: PS/PEEP , IPAP/EPAP
single
You have the option to define two different programs for therapy and
alarm settings. The program selection can be made in the Clinical
Settings menu if this option is set to dual. If dual is selected, the
patient can select between the two programs in the Treatment
screen.
Options: single, dual
cm
Selection of the unit of measure for patient height for iVAPS setup.
Options: cm/inches
-
You can erase the following data that is logged during therapy.
Used hours (therapy run hours)
High resolution data (patient flow, therapy pressure, oximetry data)
Breath-by-breath data (eg, pressure and flow related data)
Summary data (statistics in the Summary menu)
Alarm events
Settings changes
System events (eg, learn circuit passed/cancelled).
When erasing the data you will be prompted to confirm the action
before the data is erased.
Note: Motor run hours, Power-on hours and all settings cannot be
reset with this option.
during therapy, the Confirm Stop