Parameter
Description
Learn Circuit
Refer to illustration O.
The device is calibrated according to your air tubing system.
1. Ensure that the therapy is turned off before performing a Learn Circuit.
2. If in use, turn off the oxygen flow.
3. Select the mask type.
4. Set up the air circuit including accessories and patient interface.
Note: When performing a Learn Circuit for invasive use, do not connect a catheter
mount, tracheostomy tube or HMEF. Refer to Setting up for invasive use (see page 4).
5. Leave the air circuit unobstructed and open to the air.
6. Press
7. Wait for the device to complete its automated tests (<30 sec).
The results are displayed when complete. If the circuit configuration has been
successfully learnt,
(see page 19)).
FiO
Sensor
The device starts the calibration of the FiO
2
Calibration
concentration of the breathable air.
1. Press
2. Wait for the device to complete its calibration.
The results are displayed when complete.
Note: Turn off the oxygen flow.
Ramp Time
If Max Ramp Time has been set by the clinician, you may select any value up to this time.
Options: 0 min-Max Ramp Time (max 45 minutes, 5 minutes increments)

Setup menu: Configuration Menu

Parameter
Language
Brightness
Backlight
Time format
to start the Learn Circuit.
displays. If unsuccessful,
to start the FiO
2
Description
Sets the display language.
Options: Depending on regional configuration
Sets the LCD backlight brightness.
Options: 20-100%, 10% increments
Enables the LCD and keypad backlight.
If the AUTO setting is selected the backlight turns dark after five minutes without
any actions and turns on again if any button is pressed again or an alarm occurs.
Options: On, Auto
Sets the time format.
Options: 24 hrs, 12 hrs
monitoring sensor to measure the oxygen
2
sensor calibration.
displays (refer to Troubleshooting