Emerson Rosemount 3051 Series Quick Start Manual page 23

Flow meter
Hide thumbs Also See for Rosemount 3051 Series:
Table of Contents

Advertisement

August 2023
Procedure
1. Loosen the certifications label screw and slide the label to
expose the zero adjustment buttons.
2. Set the 4 mA point by pressing the zero button for two
seconds. Verify the output is 4 mA.
The optional LCD display will show ZERO PASS.
Figure 7-1: Zero Adjustment or Quick Service Buttons
Quick Start Guide
Quick Start Guide
23