Xylem wtw pHotoFlex pH Quick Start Manual page 20

Led filter photometer with integrated ph function
Table of Contents

Advertisement

Operation
No stable measured
Obviously incorrect
measured values
20
Cause
– Not enough electrolyte
– Electrode obsolete
– Electrode broken
– Socket damp
Calibration procedure:
– Incorrect solution temperature
(without temperature sensor)
– Incorrect buffer solutions
– Buffer solutions too old
Cause
value
pH electrode:
– Junction contaminated
– Membrane contaminated
Test sample:
– pH value not stable
– Temperature not stable
Electrode + test sample:
– Conductivity too low
(e.g. in ultrapure water)
– Temperature too high
– Organic liquids
Cause
pH electrode:
– Not connected
pHotoFlex
Remedy
– Top up electrolyte
– Replace electrode
– Replace electrode
– Dry socket
– Set up correct temperature
– Select buffer solutions
suitable for the calibration
procedure
– Use only once.
Note the shelf life
Remedy
– Clean junction
– Clean membrane
– Measure with air excluded if
necessary
– Temper if necessary
– Use suitable electrode
– Use suitable electrode
– Use suitable electrode
Remedy
– Connect electrode
ba75977d04
®