Xylem wtw pHotoFlex pH Quick Start Manual page 13

Led filter photometer with integrated ph function
Table of Contents

Advertisement

®
pHotoFlex
pH
Blank value
(reagent blank value)
Zero adjustment
Zero adjustment / blank
value measurement
ba75977d04
11/2018
5
Start the measurement with <START/ENTER>.
Measurement is started. The result is displayed.
Photometry \ Concentration
0.74
mg/l
1: A5/25 MC
16 mm
0.20 - 6.51 mg/l
01.02.04 15:12
A blank value is required for every concentration measurement.
For some programs (methods) for concentration measurement, the
blank values are already stored in the meter. For all other programs, the
blank value has to be determined separately before the first
measurement.
You will find more information on blank values in the photometry
analysis manual. A table with the programs and required blank values
can be found in the analysis specifications.
The zero adjustment, i. e. measuring and storing the absorbance of a
cell filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the
ambient temperature has changed.
Only perform the zero adjustment against distilled water in an optically
perfect cell. The zero adjustment must be performed separately for
each cell type.
1
Press the <M> key (long pressure) repeatedly until the
Photometry measuring mode is selected.
2
Press the <M> key (short pressure) repeatedly until the
measured parameter, Concentration is selected.
3
Press the <CAL/ZERO> key.
The menu for adjustment measurements opens up.
Photometry \ Adjustment
Zero adjust.
Measure blank value
Delete blank value
A blank value measured by
[BV]
the user is used
NH4-N
Operation
13