  • Page 3: General Information

    ® SensoLyt 900(-P) General information General information Automatic sensor The sensor electronics with the stored sensor data are in the connecting recognition head of the electrode. The data include, among other things, the sensor type and series number. In addition, the calibration data are stored in the sensor with each calibration and the calibration history is recorded (the last 10 cali- brations).
  • Page 4 ® Commissioning, measuring, calibration SensoLyt 900(-P) Commissioning, measuring, calibration Opening and closing the IDS plug connection ® This section only applies to IDS plug variants (SenTix ... -P). Plug connection closed and locked Guidance Groove Opening the plug  If necessary, clean the plug connection connection ...
  • Page 5 ® SensoLyt 900(-P) Commissioning, measuring, calibration Commissioning ® Scope of delivery  ORP electrode SensoLyt 900(-P)  Operating manual Commissioning Prepare the electrode for measuring as follows:  Remove the watering cap from the electrode tip. Possible salt deposits in the area of the watering cap do not affect the measuring characteristics and can easily be removed with deionized water.
  • Page 6 ® Storage SensoLyt 900(-P) measurement to the next by taking the following measures: – Shortly rinse the calibration and sample beakers with the solution the beakers are to be filled with next. – Between measurements, rinse the electrode with the solution that fol- lows.
  • Page 7: Maintenance And Cleaning

    ® SensoLyt 900(-P) Aging During longer storing periods, salt sediments may develop on the watering cap. They do not affect the measuring characteristics and can easily be removed with deionized water when the elec- trode is put into operation again. Aging pH electrodes are consumables.
  • Page 8: Technical Data

    ® Technical data SensoLyt 900(-P) Technical data General features Reference electrolyte Polymer electrolyte, AgCl free Junction 1-hole junction Shunt conduction element Ag/AgCl Temperature sensor Integrated NTC 30 (30 kat 25 °C / 77 °F) Measurement and pH measuring range 0.000 ... 12.000 application Allowed temperature range 0 ...
  • Page 9: Wear Parts And Accessories

    ® SensoLyt 900(-P) Wear parts and accessories Storage With watering cap; filled with KCl 3 mol/L, Ag free Disposal Residual waste Wear parts and accessories Maintenance Description Model Order no. equipment Reference electrolyte solution 250 ml KCl-250 109 705 to fill the watering cap (KCl 3 mol/l, Ag -free) Connection...
  • Page 10 ® Wear parts and accessories SensoLyt 900(-P) ba75992s02 02/2017...