pHotoFlex
pH
Measuring the ORP
Calibration
ba75977d04
11/2018
1
Submerse the ORP electrode in the sample.
pH & ORP \ ORP
157
25.0
[AR]
01.02.04 15:12
2
Press the <M> key (short pressure) repeatedly until the
measured parameter, pH is selected.
1
Press the <M> key (long pressure) repeatedly until the pH
measuring mode is selected.
2
Press the <CAL/ZERO> key.
The menu-guided calibration begins.
Follow the instructions on the display.
Calibrate
 at regular intervals
 after connecting another electrode
 when the sensor symbol flashes:
– after the calibration interval has expired
– after voltage interruption (e.g. empty batteries, empty
rechargeable battery)
mV
°C
Operation
15