Appendix: Installing EMI Controls

To ensure specified EMC performance, HDMI cable ferrites and PXI EMC filler panels must be
properly installed on your NI 5762R system. Your kit includes two HDMI cable ferrites, but PXI EMC
filler panels (National Instruments part number 778700-01) must be purchased separately. For more
installation information, refer to the NI FlexRIO FPGA Module Installation Guide and Specifications.
Installing PXI EMC Filler Panels
Complete the following instructions to install PXI EMC filler panels (National Instruments part
number 778700-01) in your PXI chassis:
1.
Remove the captive screw covers.
2.
Install the PXI EMC filler panels by securing the captive mounting screws to the chassis, as shown
in Figure 15. Make sure that the EMC gasket is on the right side of the PXI EMC filler panel.
1
2
1
Captive Screw Covers
Note
You must populate all slots with a module or a PXI EMC filler panel to ensure proper module
cooling. Do not over tighten screws (2.5 lb · in. maximum). For additional information about the use
of PXI EMC filler panels in your PXI system, visit
© National Instruments Corporation
3
1
2
Captive Mounting Screws
Figure 15. PXI EMC Filler Panels and Chassis
25
3
EMC Gasket
and enter
ni.com/info
NI 5762R User Guide and Specifications
.
emcpanels