DIGITAL CALIBRATION
Solder Iron Digital Temperature Calibration Example
The external temperature sensor displays 248 to 252 degrees.
●
The set temperature and displayed actual temperature of the
●
soldering iron is 300 degrees.
300 — 248 = 52. An additional adjustment of 52 degrees is
●
required.
Enter calibration mode
●
We increase from "0000" to "0052".
●
Exit calibration mode.
●
The external temperature sensor would now display 298 to 302
●
degrees.
NOTES:
The calibrated data is saved into the memory an shall remain
●
effective until it is recalibrated again or new data is entered.
Calibration will only make the newly calibrated point the most
●
accurate. Other temperature points may be a little off.
The soldering Iron has a lowest temperature limit such that when
●
the temperature has been set to 200 degrees with external actual
temperature also showing 200 degrees, further decreasing the
temperature offset would be only have
actual temperature.
To reset the calibration settings to factory setting, switch
●
OFF and ON the unit and while holding the hot-air temperature
up
button
temperature / Temperature calibration of the unit would revert to
its default factory setting.
until
the
banner
minimal effect to the
finishes
scrolling,
the
Set
15