Xylem SL Analytics Lab 865 Operating Manual page 32

Ph meter
Hide thumbs Also See for SL Analytics Lab 865:
Table of Contents

Advertisement

pH value
Calibration record
(example)
32
Display
Clean the electrode according to the
electrode operating manual
Error
Error elimination (see section 11 W
..., page 55)
TO DO IF
Lab 865
Ser. no. 11292113
CALIBRATION pH
01.02.2014 15:55
Ser. no. 10501234
Buffer 1
Buffer 2
Buffer 3
Voltage 1
Voltage 2
Voltage 3
Temperatur 1
Temperatur 2
Temperatur 3
Slope
Asymmetry
Sensor
etc...
Calibration
Zero point
record
[mV]
-
- 30 ... <- 25
or
->+25 ... + 30
Error
- 30 ... + 30
HAT
7.00
184.0 mV
3.0 mV
-177.0 mV
24.0 °C
24.0 °C
24.0 °C
-60.2 mV/pH
4.0 mV
+++
Lab 865
Slope
[mV/pH]
-62,0 ... <-61,0
or
>-56,0 ... -50,0
-62,0 ... -50,0
ba77145e02
12/2017