Xylem SL Analytics Lab 865 Operating Manual page 29

Ph meter
Hide thumbs Also See for SL Analytics Lab 865:
Table of Contents

Advertisement

Lab 865
Continuing with
three- to five-point
calibration
ba77145e02
12/2017
17.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value appears.
pH
4.035
ConCal
01.02.2014
18.
Set the nominal buffer value for the measured temperature with
<><>.
19.
Accept the set calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
20.
Finish the calibration procedure as a two-point calibration with <M>.
The calibration record is displayed.
21.
Thoroughly rinse the electrode with deionized water.
22.
Immerse the electrode in the next buffer solution.
23.
When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
24.
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the setting of the nominal buffer value appears.
pH
9.958
ConCal
01.02.2014
2
Buffer

24.8
°C
08:00
3
Buffer

24.8
°C
08:00
pH value
29