4.0  BASIC
OPERATION 
 
4.1  TURNING
ON
 
Plug in the unit using the DC adapter or use the internal rechargeable battery. It is 
recommended that you charge the battery for at least 8 hours before first use.  
1) To turn on press the [on/off] key once and release. The balance will show the 
capacity and the software revision, and then self‐test before showing zero on 
the display and the stable sign. 
2) The balance is now ready to be used.  
3) To  turn  the  balance  off  after  use,  press  the  [on/off]  key  again.  There  is  an 
auto power‐off function that will automatically turn the unit off if not used for 
a period of time, this can be set in the parameters section. 
If  an  error  is  shown  while  turning  on  the  balance  check  that  the  pan  is  fitted 
correctly  and  that  there  is  no  weight  on  the  pan  or  the  internal  calibration  weight 
(not approved models) has not been left in the calibrate position. Move the lever all 
the way to the left.   Turn off the balance and turn it back on again. 
4.2  ZEROING
/
 
You can press the [‐>0/T<‐] key to set a new zero point and show the zero reading if 
the  weight  reading  is  less  than  4%  of  the  total  of  the  maximum  capacity  of  the 
balance (+/‐ 2% for approved models). This may be necessary if the weight reading is 
not  reading  zero  with  nothing  on  the  pan.  The  zero  indicator  will  show  up  in  the 
upper left corner of the LCD.  
If  you  are  using  a  container  to  weigh  then  you  can  place  this  on  the  platform  and 
press  the  [‐>0/T<‐]  key.  Providing  the  container  weight  is  more  than  4%  of  the 
maximum capacity of the balance (+/‐ 2% for approved models), the digits will show 
zero  and  NET  will  light  up  on  the  display.  You  can  then  weigh  your  sample  in  the 
container. Taring weight subtracts from the total balance capacity.  
 
 
 